Tag Archives: a/w syringe button

  • Tech Tips #51: Cross-Over Part 2

    Last month in Tech Tips we discussed a few possible sources of cross over and how to determine if you actually have cross over. We finished up by looking at master valves.

    What if you’ve got cross over in more than one room? Is it possible that one room is affecting the others? Yes, cross over can work through the lines from one room to the entire office. In fact, you can even get air in your pipes (if you’ve got air in the water line) so pipes will rattle, squeak, or groan when any source of water is used (such as a faucet or even a toilet). Of course, you can also get failed valves in multiple rooms at once.

    How do you determine which room is to blame, or even tell if it’s one or more?

    Since we left off last month in the junction box, let’s continue there.

    In addition to your master valves, you should also have manual shut offs in your junction box. Normally, the master valves are attached directly to the manual shut-offs. The manual shut-offs have a knob that you can turn to physically stop the flow of air and water. Use them to shut both air and water off in an affected room. Start in the room closest to the compressor.

    A

    Once the manual shut-offs are closed, depress the buttons on your air/water syringe until all flow ceases. Next turn the AIR manual valve back on, but LEAVE THE WATER OFF. Put your syringe tip into a cup of water and hit the WATER button of the syringe. If you have bubbles, you have cross over in that room. If you had been experiencing water in the air line, you’ll want to hold the syringe button down for at least one full minute as it may take some time for the air to work its way through the lines. If you don’t have bubbles, that room is probably fine. Turn both air and water off with the manual shut-offs and proceed to the next room and repeat this test. It’s important that you turn both valves off before checking the next room to keep any cross-over that might be occurring in this room from affecting others.

    Where else can you get cross over? Any other air-activated water valve in the unit.

    If you have cuspidors with timers for the bowl rinse or cup filler functions, the timers are air activated and could be a source of cross over. As cuspidors typically have large lines attached, these can affect the whole office more readily. If you’ve got squeaking pipes, your cuspidor may be a more likely culprit.

    Just like master valves, timer valves can vary in design but will be mounted inside some sort of “box” or “utility center” directly below the cuspidor. The timers often have a similar appearance to master valves and will have some sort of knob to adjust the timing.

    The

    Other possible sources of cross over are the water relay valve or even the handpiece block.

    Normally, a failure in the water relay valve will result in trouble with handpiece water long before you experience cross over, so this is not very likely unless you’ve also had trouble with your handpiece water coolant (dripping, coming out of the wrong handpiece, or other issues). Depending on the design of valve, usually the water relay just needs to be replaced as most are not readily serviceable.

    Last of all is the handpiece block. There are many designs of block, so this can be more difficult to check, but often a diaphragm will be readily apparent when you inspect the block (such as in the block pictured below, as used in a Beaverstate delivery system). You can clearly see the seam between the two layers of brass in the photo. This is where the diaphragm is.

    A

    In many models of units, water does not actually enter the handpiece block, so you would not get cross over here. Always be sure to verify the configuration of your unit before checking the block.

    If you have quick disconnect handpieces, you can also get cross over in the coupler. Normally cross over here would only be seen in the handpieces. In this case, you may see water dripping out of the unit. This is actually from water coming out the exhaust line (which often empties into the unit). If you see water dripping out of the unit (or it’s getting into your turbines) change the o’rings on your quick disconnect.

    As you can see, there are a number of possible sources of cross-over in the dental office. To help you diagnose, we’ve provided a handy checklist below:

    Possible Sources of Cross-over

    • Compressor
    • Air/water syringe
    • Master Valves
    • Cuspidor timer blocks (if present)
    • Handpiece block
    • Handpiece coupler (if a quick disconnect)
    • Any other air activated valves (e.g. water relay)

    As mentioned above, check one room at a time to verify where cross-over is occurring then check the valves within each room to determine the precise source.

    Save big with our End-of-Year Specials! Hot deals on gloves, diamonds, carbides, & dental equipment. Prices valid through 12/31/12.

  • Tech Tips #50: Cross-Over- Part 1

    What is cross-over? What causes it, and how do you fix it?

    And no, we don’t mean a 4 wheel drive vehicle that combines the features of a sedan, wagon, and SUV.

    As we discussed way back in Tech Tips #3 Dental Pneumatics, most dental units run on air and require air for just about everything, including turning on the water (which is “air activated”). This means there are valves inside the unit which have both air and water flowing to (and possibly through) them. A failure in one of these valves can lead to cross-over, getting the air in the water line or water in the air line.

    The first step is to verify that cross-over is actually occurring. If you have water in the air line, start by checking at the compressor. Do you have good dry air coming out of the compressor? If the air is moist right at the compressor, the problem is at the compressor itself and you don’t have cross-over. Check the filters, dryer, and drain the tank. If the problem persists, additional trouble-shooting of your compressor may be required, call and speak with one of our technicians.

    If the air is dry at the compressor, then you probably have cross-over, so now you need to figure out where it’s happening.

    Are you experiencing cross-over in just one room or several? If it’s in just one room, that will narrow things down considerably; if it’s in more than one, you could have multiple failures, or you could have water or air working its way through the lines from one room to the others. Either way, we’ll next want to check within one room that’s affected.

    Are you getting cross-over in the air/water syringe? The syringe air and water doesn’t typically go through many valves, so if you’ve got cross-over here, it’s likely from a failure within the syringe itself, or in the master valves (see Tech Tips #23 ”Getting to Know Your Utility Center”).

    If cross over is occurring in the syringe itself, the cross-over could be isolated to the syringe. Check for cross-over elsewhere. Do you have trouble with the handpiece drive air? If you’'ve got water in the drive air, you’ll often notice that water drips out the bottom of the unit when you use your handpiece (as the drive air typically exhausts into the unit). Do you have an assistant’s side syringe as well? Is it affected? If you don’t have cross-over anywhere but in one syringe, then the cross-over is quite possibly within that syringe.

    If it’s just one syringe, the most likely culprit is the small internal o’ring, our part #01-06. If this o’ring is missing or damaged, you can get cross-over in your syringe.

    If the water just fails to shut off, you’ve probably got a worn button or valve core. Check our previous issues on syringe repair and syringe troubleshooting for more information on resolving these issues.

    Another possible point of cross-over in the syringe is unique to certain brands of autoclavable syringe (such as A-dec). With these syringes, a manifold is inside the handle and the syringe head attaches to this manifold and is secured in place when the handle is threaded on. The manifold has two male connectors that are off-center and which correspond to two holes on the bottom of the syringe. One port is for air, the other for water. These ports are off-center so they will only connect to the syringe if properly aligned. In theory, this will prevent cross-over, however in reality it can cause it. If the ports are not properly aligned, the handle can still screw onto the head of the syringe, but the air and water will just flow into the cavity between the manifold and the syringe head. This creates a constant mix in the syringe handle which will yield cross over in the syringe. Ironically, even though this occurs at the syringe, this particular source can lead to air or water flowing back down the lines and yielding cross over in other items in the operatory as well. If you are getting cross over and have autoclavable syringes, always make certain that the syringe head is securely attached to the manifold before proceeding with other checks (if for no other reason, this is a very quick and simple thing to check).

    This photo shows an A-dec syringe handle with the manifold.

    If the syringe checks out, then we need to check other possible sources of cross over.

    Every air activated valve in the unit is a potential source of cross-over, and all of them should be checked.

    As mentioned previously, if you’re getting cross-over in the syringe, you could also have a failure in your master valve. The master valve is one of the most common sources of cross-over we encounter. There is a diaphragm on the master valves that corresponds to the air activation. Usually the diaphragm is under a square cover on the valve. This will also be where an air line attaches to the valve.

    A few common designs of master valve are shown below with the location of the diaphragm pointed out:

    Note the square covers held on with screws at the corners in the second and third photos. This type of design is very common on master valves. Look for a square cover on your master to find the diaphragm.

    The third photo is of an A-dec water master. Note the yellow tubing with a red dash. This is the A-dec color-code for an air activation line (air out from the master switch on the front of the unit). The diaphragm will usually be near where this line attaches (another thing to look for to correctly identify where your diaphragm is). American Dental Accessories, Inc. carries master valve repair kits which will allow you to replace the diaphragm. In addition, the repair kits include o’rings, valve stems and springs. We recommend replacing all of these components if you’re accessing the valve interior. You can also see a manual shut-off in the third photo. The master valve normally attaches directly to the manual shut-off. You will want to use the manual shut-offs to turn the air and water off at the unit before servicing the master valve(s).

    It is also common to have a relief hole in the side of the master valve just above the diaphragm. If water is coming out of this hole, this is another indication of a ruptured diaphragm.

    A worn or damaged diaphragm in the water master can also lead to getting air in your water line.

    Stay tuned for Part 2 next month. We’ll cover more valves that could be the source of cross-over, details on how to narrow down which operatory might be the source of cross-over, and the effects of air in the water line. All of this and more, next month!

  • Tech Tips #43: Air/Water Syringe Troubleshooting

    Air/water syringes are the workhorses of the operatory but can often exhibit minor malfunctions. Most difficulties you encounter with your syringe can be easily addressed in just a few minutes by the office staff.

    Air mixing with water (or water mixing with air).

    If either button is sticking, that could cause the air or water to continue running and lead to this problem. See “sticking buttons” below. If the buttons are not sticking, the problem is likely inside the syringe head.

    There is a small o’ring inside the syringe which keeps the air and water separate (the “small adaptor nut” o’ring part 01-06). This o’ring could be worn or missing and should be replaced.

    It is also possible that there is condensation in the air line from the compressor. In this case, you would only have trouble with water mixed in when hitting the “air” button, and not air mixed in when hitting the “water” button. An easy way to determine if you have water in your air line from outside the syringe, is to detach the air line from the bottom of the syringe, and let the air from the line blow into a paper towel or onto your hand. You’ll know pretty quickly if there’s moisture in the line.

    Water can also get into the air line from other valves in the unit. See some of our previous issues for more information on trouble shooting water in the air line from other sources.

    No Air

    No air comes out when hitting the air button. This is almost always due to an obstruction somewhere. Start by trying without a tip installed on the syringe. If you now get air, the obstruction was in the tip. Putting it in an ultrasonic cleaner for a minute or two may clear it up. Otherwise, discard it as the tip is worn out.

    If you get no air without a tip either, the clog is further upstream. Remove the holding mechanism from the syringe and check for debris on the inside of the head. Usually the holding mechanism is unscrewed using a 5/32” allen (hex) wrench. Remove the button and valve core (if your syringe uses valve cores) and check for debris here as well.

    Detach the tubing from the base of the syringe and verify you have air pressure in the line. If you do not have air pressure in the tubing, the clog is at a valve or junction before the syringe. Call our techs to trouble-shoot further (or see our previous issues on leaking, junction boxes, & pneumatics).

    No water

    Follow the same procedure as above for no air. Additionally, if you use plastic barriers over your syringe, it is very common to get debris on the inside as staff will often poke the syringe tip through the barrier when changing tips. If the tip is poked through the barrier, every time the tip is poked through, a small piece of plastic is shoved into the syringe with it. We frequently see syringes with a huge wad of plastic built-up inside the small adaptor nut o’ring (#01-06, see above). Check carefully as this is typically clear plastic and hard to see. You should be able to remove it with a scaler or explorer. You might also want to show your staff to make certain they cease using this technique.

    If there is no debris evident inside the head, continue as above (under “no air”).

    Sticking Buttons

    This is more frequently experienced with the water button than the air button. Mineral deposits from hard water, debris in the water line, or corrosion can all wear the water button and the o’rings on it. If it is sticking, remove the button and lubricate with a silicone-based lubricant. In fact, it’s a good idea to regularly lubricate your buttons as part of  routine maintenance to keep them moving smoothly. We recommend once a month, but you may want to adjust the frequency depending on your conditions and experience (e.g. if you have particularly hard water, a greater frequency may be warranted).

    Lubrication may be enough to restore proper function. If not, you will need to replace the button (or valve core, if your syringe uses valve cores). As a temporary fix, you can swap the air and water buttons of your syringe. This will cause the air button to stick, but this is often less problematic than having the water constantly leak.

    See our previous issues on  syringe repair for step-by-step instructions on removing and changing buttons.

    Spinning Tips

    Most syringes are designed for quick changing of the tips so the tips are just held in with o’rings. This allows the tip to be easily pulled out or pushed in- no loosening or tightening of components. Older designs used a brass or plastic syringe cone to help retain the tip, but these required use of a wrench to unscrew or tighten the holding mechanism to change tips. The cones, however, often had a slit that ran the length of the cone which would restrict lateral movement of the tip.

    Newer designs that just use round o’rings do not provide the same lateral tension. You’ve got a round peg in a round hole. Tightening the holding mechanism can compress the o’rings further limiting rotation of the tip, but it’s difficult to tighten enough without preventing easy tip changes.

    Most current designs use two o’rings to hold the tip; the first o’ring (closest to the tip) can be replaced with a cone to limit rotation. Not all assemblies have room for a cone here, but most should.

    The DCI collar-type syringe uses ball-bearings to hold the tip in place. DCI makes a tip with extra notches on the end of the tip. These notches will fit in-between the ball bearings of their collar mechanism making it harder to spin the tip. The DCI collar has 6 ball bearings in the mechanism, some of the other brands of collar syringe use fewer ball bearings so the notches won’t limit rotation as effectively in these syringes.

    Last of all, you can rely on simple physics. Rather than pushing on the side of the tip for retraction, use the hook-like end of the tip to pull. Syringe tips are designed to stay in place when pulled on with only modest force, tips are not designed to resist rotational forces.

    Leaking- air or water not shutting off

    Most syringes currently on the market have a spool-type button that serves as the valve. The button itself turns the air or water on. If the water or air is staying on, typically the button has failed. If the button has failed, simply replace it. A sticking button will also lead to failure of air and water to shut-off. See above under “sticking buttons” for possible remedies.

    Some (mostly older) syringes will use a valve core under the button. This is what actually turns the air and water off and on, the button simply activates the valve core. If your syringe uses valve cores, typically the valve core will need to be replaced using a valve core wrench.

    Leaking- air or water coming out somewhere other than the tip

    “Leaking” can also mean air or water coming out from somewhere other than where it is supposed to. Most components of the syringe are sealed with o’rings. Leakage around any of these components will usually be remedied by replacing the corresponding o’ring.

    For leakage around the holding nut, replace the large skinny o’ring (#01-05) that seals it. If you have leakage out around the buttons, replace the buttons (unfortunately, o’rings are not available for most buttons, so the entire assembly should be replaced).

    A note on disposable tips:

    Many disposable tips on the market require a special adaptor to properly fit your syringe. If not using this adaptor, many problems can be observed such as water mixing in with the air (see above), poor spray pattern, or even tip ejection (which can cause harm to patients). If your disposable tip does not have the same shape and design as a standard tip (a skinny water line protruding out the back and a groove toward the back to line up with the o’rings) it will require a special adaptor. Regardless of brand or style, it's always a good idea to give the tip a little tug after inserting to make certain that it's properly and securely seated.

    However, if the tip is identical to an autoclavable tip (just made of plastic in whole or in part – such as our Smart Tips), then you probably don’t need to use an adaptor. Of course, if you use autoclavable tips you’ll avoid this issue altogether (to say nothing of reducing waste and saving money).

    For step-by-step guidelines are performing the various repairs mentioned above, check out previous issues of Tech Tips:
    How to Repair a Leaky Syringe- Part 1
    How to Repair a Leaky Syringe- Part 2
    How to Rebuild Your Air/Water Syringe- Video